VT Plus HF

Operators Manual

2-6

3. Ensure that the breath detection mode is set to Unidirectional Expiratory.

a. Go into

Setup

b. Choose

Settings

c. Choose

Breath Detect

and modify as needed

The Analyzer displays a screen showing the flow of gas delivered by the ventilator:

fec049.bmp

The instantaneous flow rate displays in large numbers in the upper right portion of the

screen. The flow statistics (min., max and average) are shown below in the center right

portion of the screen. A subset of the breath parameters is shown below the flow plot.
The following adjustments can be made to the screen:

Switch between the high- and low-flow ports – Press the

RANGE

soft key.

Change the measurement units (LPM, CFM, LPS) – Press the

UNITS

soft key.

Change the plot scaling – Press the

RESCALE

soft key. The new scaling is based on

the minimum and maximum values shown in the statistics area.

Erase the plot and re-initialize the statistics and breath parameters – Press the

CLEAR

soft key.

See more information in any screen – Press the

HELP

key.

Pause and resume the plot screen, respectfully – Press the

PAUSE/RESUME

key.

Printer Cable Attachment

The printer uses a D25 male to Centronics parallel cable for parallel printing. This cable

is available from Fluke (part #2238072) or from most electronic supply outlets.
To attach the parallel cable:
1. Connect the 36-pin end for the Centronics type parallel cable to the printer’s parallel

input connector.

2. Attach the 25-pin male end of the parallel cable to the 25-pin connector on the back

of the Analyzer.