Operators Manual

3-18

Full Breath Parameters Test Screen

The Full Test screen displays all of the breath parameters calculated by the Analyzer.

fec055.bmp

Note

Breath detection must be enabled for breath parameters to be calculated.

See System Setup for instructions on enabling breath detection.

Trend Press

the

TREND

soft key to print the trend of a single parameter. See

Special Functions for a more description of the trend function

Clear Press

the

CLEAR

soft key to reset the statistics and re-start the plot.

If asterisks appear for a parameter value, the parameter is either invalid for the current

breath detect mode or the system does not have sufficient information to accurately

calculate the parameter. For example, Base Flow is invalid during bi-directional breath

detection mode because the base flow is seen only in the inspiratory and expiratory limbs

of the ventilator circuit. In another example, it would be impossible to measure the

inspiratory hold pressure during the expiratory-only mode.

Note

The Assist Pressure (Assist) and the Minimum Pressure (Min Pres) are not

available for trending and cannot be added to other screens using the

parameter selection. Refer to Special Functions for instructions on Assist

Mode testing.