B. error messages, Error messages, Appendix b – Fluke Biomedical VT Plus HF User Manual

Page 133

Advertising
background image

B-1

Appendix B

Error Messages

Table B-1 lists error messages, their descriptions, and corrective actions.

Table B-1. Error Messages

Error Message

Description

Corrective Action

Flow Out of Range

Differential pressure for the flow
measurement causes a
measurement greater than the
maximum (or less than the
minimum) reading of the analog
to digital converter.

Ensure that flow is within the
specified range of the Analyzer.

Gas Settings Error

Oxygen concentration is
measured at greater than 50 %
and the gas settings are not set
to an oxygen or a balance
oxygen setting. This is provided
as a warning to set the gas
settings if measuring flow with a
gas type that has high oxygen
concentrations.

The error/warning occurs only
once after the Analyzer is
powered on. If measuring flow,
ensure that the correct gas
setting is being used for gas type
being measured.