C. troubleshooting, Troubleshooting, Appendix c – Fluke Biomedical VT Plus HF User Manual

Page 137

Advertising
background image

C-1

Appendix C

Troubleshooting

Table C-1 lists problems, possible causes, and corrective actions.

Table C-1. Troubleshooting

Problem

Possible Cause

Corrective Action

Analyzer does not turn on –
Screen is dark

No power Screen contrast not
set Blown fuse

Plug unit into an AC mains outlet
using the supplied power cord.
Turn the switch on. Adjust
screen contrast using multiple
pushes of the contrast key on
the side of the Analyzer display.
Replace fuse with one of correct
rating as marked on the
Analyzer rear panel or as
specified in the manual.

Analyzer screen turns on but is
just white. Splash may or may
not appear before screen goes
white

Flash program memory not
loaded Flash memory or boot
memory resulted in incorrect
checksum. Screen contrast not
set

If problem persists, call Service.
If problem persists, call Service.
Adjust screen contrast using
multiple pushes of the contrast
key on the side of the Analyzer
display.

Screen is difficult to read

Screen contrast not set

Adjust screen contrast using
multiple pushes of the contrast
key on the side of the Analyzer
display.

Flicker is observed on left side of
Analyzer screen

Slight flicker is normal

Adjust contrast to diminish
effect.