Introduction and Specifications

Specifications

1

1-13

Table 1-7. Breath Parameter Accuracy Specifications

Parameter

Resolution

Range

Accuracy

Inspiratory and
Expiratory Tidal Volume

0.1 ml

As specified in

high/low-flow spec

As specified in

high/low-flow spec

Minute Volume

0.001 lpm

0-60 l

3 %

Breath Rate

0.1 bpm

0.5 – 150 bpm

1 %

Inspiratory to Expiratory
Time Ratio (I:E Ratio)

0.01

1:200 to 200:1

2 % or .1

Peak Inspiratory
Pressure

0.1 cmH

2

O ±120

cmH

2

O

±3 % or 1 cmH

2

O

Inspiratory Pause
Pressure

0.1 cmH

2

O ±120

cmH

2

O

±3 % or 1 cmH

2

O

Mean Airway Pressure

0.1 cmH

2

O ±80

cmH

2

O

±3 % or 0.5 cmH

2

O

Positive End-expiratory
Pressure (PEEP)

0.1 cmH

2

O

-5 to 40 cmH

2

O

±3 % or 0.5 cmH

2

O

Lung Compliance

1

0.1

ml/cmH

2

O

0 - 150 ml/cmH

2

O

±5 % or 5 ml/cmH

2

O

Inspiratory Time

0.01 sec

0 - 60 sec

0.5 % or 0.02 sec

Inspiratory Hold Time

0.01 sec

0 - 60 sec

1 % or 0.1 sec

Expiratory Time

0.01 sec

0 - 90 sec

0.5 % or 0.01 sec

Expiratory Hold Time

0.01 sec

0 - 90 sec

1 % or 0.1 sec

Peak Expiratory Flow

0.01 lpm

0 - 300 lpm

3 % or 2 lpm

Peak Inspiratory Flow

0.01 lpm

0 - 300 lpm

3 % or 2 lpm

Flow Bias

2

0.01 lpm

0 - 30 lpm

2 % or 0.5 lpm

1

Inspiratory pause time >0.5 sec

2

Expiratory pause time >0.5 sec