Performance specifications – Fluke Biomedical VT Plus HF User Manual

Page 23

Advertising
background image

Introduction and Specifications

Specifications

1

1-11

Power Consumption ....................... <132 VA
Fuse Rating .................................... 0.5 A, Slo-Blo

Performance Specifications

Low-Pressure Port

Maximum Applied Pressure............. 60 psi
Operating Pressure ......................... (Differential) ±500 mmHg (±10 psi)

(Common-mode) 30 psi

Span Accuracy ................................ ±0.80 % of reading or ±1.5 mmHg, whichever is greater
Frequency Response ...................... >10 Hz
Resolution........................................ 0.1 mmHg
Sample Rate ................................... 100 Hz

Note

Fluid pressure may be applied to the positive port; however, fluids should

be kept from entering the pressure port by using a suitable length of

connection tubing.

High-Pressure Port

Maximum Applied Pressure............ 150 psi
Operating Pressure ........................ ±100 psi
Span Accuracy ............................... ±1 % of reading or ±0.3 psig, whichever is greater
Frequency Response ..................... >10 Hz
Resolution....................................... 0.1 psi
Sample Rate................................... 100 Hz

Note

Fluid pressure may be applied to the positive port; however, fluids should

be kept from entering the pressure port by using a suitable length of

connection tubing.

Airway Pressure

Maximum Applied Pressure............. 20 psi
Operating Pressure ........................ ±120 cmH

2

O

Span Accuracy ............................... ±0.75 % of reading or r 0.5 cmH

2

O, whichever is greater

Frequency Response ..................... >25 Hz or t10-90 <40 ms, whichever is greater
Resolution....................................... 0.1 cmH

2

O

Sample rate .................................... 100 Hz

Note

Airway pressure is internally tapped off the proximal flow sensor port,

which is the port closest to the exhaust port on the Analyzer.

Continuous Flow

Low Flow ............................................... ±(2 % of reading and 1 % of range)

High Flow............................................... ±(2 % of reading and 1 % of range)

Volumetric Flow

High-Flow Port

Maximum Flow Rate........................ (absolute value) 500 lpm
Operating Flow Range.................... ±300 lpm
Accuracy......................................... ±2.0 % of range
Resolution....................................... 0.01 lpm