Breath detection – Fluke Biomedical VT Plus HF User Manual

Page 39

Advertising
background image

Connection and Setup

Setting Testing Parameters

2

2-11

Note

The Analyzer accepts barometric pressure values between 0 and 999

mmHg. No decimal values or fractions are allowed.

4. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

5. Press the

BACK

soft key to return to the

Settings

menu.

Breath Detection

The Analyzer can be used to measure parameters on a breath-to-breath basis. To measure

breath parameters, the mode of breath detection must be set. The breath detection options

are shown in Table 2-2 .

Table 2-2. Breath Detection Options

Option Description

Q

Bi-directional breath detection – Use this breath detection mode when analyzing
both inspired and expired breath data or when the Analyzer is connected
between the load (lung simulator) and the Y-piece in the breathing circuit.

N

Unidirectional inspired only – Use this breath detection mode when the Analyzer
is analyzing only inspired breaths or when the Analyzer is connected in the
inspired limb of the ventilator breathing circuit.

P

Unidirectional expired only – Use this breath detection mode when the Analyzer
is analyzing only expired breaths or when the Analyzer is connected in the
expired limb of the ventilator breathing circuit.

Off

Use this mode to measure continuous flow and volume without attempting to
detect breaths. Use this mode to integrate a flow over a long period of time.

Setting Breath Detection Mode

To select the breath-detection mode:
1. From the

Gas Settings

menu, highlight

Breath Detect

.

2. Press the

MODIFY

soft key to toggle to the desired value.

3. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

4. Press the

BACK

soft key to return to the

Settings

menu.

Setting Breath Detect Threshold

The Analyzer breath detection algorithm uses a flow rate threshold that the flow must

cross to trigger a change in the breath phase. Higher flow rate thresholds yield less

sensitive breath detection. Lower thresholds yield more sensitive breath detection.
To enter the breath detect threshold:
1. From the

Gas Settings

menu, highlight

LF

Threshold

to adjust the low-flow

threshold or

BD Threshold

to adjust the high-flow threshold.

2. Press the

MODIFY

soft key to display the value field.

3. Input the desired value from the keypad.