VT Plus HF

Operators Manual

4-20

Ending the Stacked Volume Test

To end the Stacked Volume test:
1. Either click the

STOP

button or let the Stacked Volume test end following the

number of requested breaths.

2. Click the

Print

button to print a final report for the Stacked Volume test.

3. Click the

New Test

button to start a new Stacked Volume test.

Serial Communications

The Analyzer communicates with a PC through a serial (COM) port. If the Analyzer is

connected to a computer and turned on when the software is started, the software

automatically detects which COM port the Analyzer is on and then initiates

communications between the Analyzer and your PC.

Selecting COM Port

While running VT for Windows the user may select

Com Port

from the

Communications

menu to change communications settings.

Selecting the Serial Communications Mode

The Analyzer has four modes for data transmission via the serial port, available from the

Communications

menu. The currently selected mode shows a check mark next to its

name.

Test Parameter

–the parameter that was chosen for running this test; this value does

not change during the test.

Wave Form

– starts the continuous transmission of waveform data from the Analyzer;

flow, volume, airway pressure, auxiliary pressure, and differential pressure data are

received at a 50 Hz rate.

Both

– starts the automatic transmission of both waveform data and parameter data, as

described above.

Parameters

– start the automatic transmission of parameters that are calculated at the

end of a breath; these parameters are calculated on a breath-to-breath basis.

Note

No parameters are calculated for a breath if a zero occurs during that

breath.

None

– selects no data transmission; this stops all automatic data transmissions from the

Analyzer.