Barometric Pressure

The Analyzer automatically calculates the barometric pressure to within 2 % of full scale.

To give greater accuracy, the Analyzer allows the user to calibrate the barometric

pressure offset by entering the local barometric pressure into the system. This value is

used to calibrate the system barometric pressure offset during future use.
To change the barometric pressure:
1. From the

Setup

menu, select

Barometric Pressure

.

2. Enter the local barometric pressure in the box provided.
3. Click

OK

.