Fluke Biomedical VT Plus HF User Manual

Page 98

Advertising
background image

VT Plus HF

Operators Manual

4-14

Running the Trend Test

To run the Trend test:
1. Click the

START

button. The screen changes, as shown in Figure 4-9.

fec091.bmp

Figure 4-9. Trend Test Screen: Testing Underway

While a test is running, screens cannot be changed until the

STOP

button that

appears at the bottom of the screen has been clicked.
During a trend test, the following values are reported to the user every time a breath

occurs:

Test Parameter

– This is the parameter that was chosen for running this test. This

value does not change during the test.

Current Value

– This is the current value of the chosen test parameter. Every time a

breath occurs, this is updated and compared against the initial value of the test parameter.

Starting Value

– Shows the initial value for the test parameter. This value does not

change during the test.

% Difference

– This value is calculated each time a breath occurs. If it is greater than

the % limit, an incident is reported.

% Limit

– This is the value entered before starting the test. It does not change during the

test.

# of Tests

– Incremented each time a breath is detected.

# of Incidents

– The number of incidents is incremented each time the percent

difference exceeds the percent limit.

Elapsed Time

– This value is updated continuously. It keeps track of the length of time

the current trend test has been run.