VT Plus HF

Operators Manual

4-18

Ending the Leak Test

To end the Leak test:
1. Either click the

STOP

button or let the Leak test end by the time running out. When

the Leak test is over a report is prepared showing the final values for the calculated

parameters.

2. Click the

Print

button to print a final report for the Leak test.

3. Click the

New Test

button to start a new Leak test.

Stacked Volume Test Screen

The Stacked Volume test measures tidal volumes over a given number of breaths and

calculates the average tidal volume.

Setting Stacked Volume Test Parameters

The Stacked Volume test setup screen, shown in Figure 4-13, allows the user to enter the

number of breaths to be averaged.

fec095.bmp

Figure 4-13. Stacked Volume Test: Setup Screen