Operation

Special Functions

3

3-25

7. Use the keypad to enter the duration of the Leak test in hours, minutes, and seconds.

Six digits must be entered to set the test time. The time is entered in the following

format: HHMMSS (HH is the hours (0-99), MM is the minutes (0-59) and SS is the

seconds (0-59). Be sure to use leading zeros to enter values less than 10. For

example, to enter a time of 1:09:05, press the following sequence of keys: 0, 1, 0, 9,

0, 5).

To run the test:
1. After all of the test parameters are set, pressurize the test chamber (test lung).
2. Press the

START

soft key

to begin testing. A screen similar to the one shown is

displayed.

fec062.bmp

3. The display shows a pressure plot and instantaneous pressure value to the right of the

plot. Parameters below the pressure plot show the entered compliance, the time of

day that the test was started, and the calculated test results.

4. Press the

STOP

soft key to terminate the test or let the test time elapse. In either

case, a Leak test result displays.

fec063.bmp