Breath parameter accuracy specifications – Fluke Biomedical VT Plus HF User Manual

Page 24

Advertising
background image

VT Plus HF

Operators Manual

1-12

Frequency Response. .................... >25 Hz or t10-90 <40 ms, whichever is greater
Sample Rate................................... 100 Hz
Dynamic Resistance....................... <2.00 cmH

2

O @ 60 lpm

Low-flow Dropout............................ 25 lpm
Breath Detect Threshold................. (user settable) 2.0 lpm
Volume Range................................ >±60 l

Notes

Tidal Volume Accuracy ±3 % of reading or ±10 ml, whichever is

greater

Volume Accuracy tested to 7 liters.

Flow accuracy is specified for dry air or oxygen.

Do not use with humidified gases.

Low Flow Port

Maximum Flow Rate........................ (absolute value) 50 lpm
Operating Flow Range.................... ±25 lpm
Accuracy......................................... ±1.0 % of range
Resolution....................................... 0.01 lpm flow >1 lpm; 0.001 lpm flow <1 lpm
Frequency Response ..................... . >25 Hz or t10-90 <40 ms, whichever is greater
Sample Rate................................... 100 Hz
Dynamic Resistance....................... <2.5 cmH

2

O @ 5 lpm

Low-flow Dropout............................ 0.01 lpm
Breath Detect Threshold................. (user settable) 0.50 lpm
Volume Range ............................... >±60 l

Note

Tidal Volume Accuracy ± 3 % of reading or ±5 ml, whichever is

greater

Volume Accuracy tested to 1 liter

Flow accuracy is specified for dry air or oxygen.

Do not use with humidified gases.

Below 0.25 lpm, measurement accuracy is obtained by allowing the

Analyzer to fully warm up, and manually zeroing before reading or

documenting a measurement.

Breath Parameter Accuracy Specifications

Breath parameter accuracy specifications are listed in Table 1-7.

Note

If a choice of readings is indicated by an “or”, choose whichever is

greater; for example, 3 % or 250 ml.