Inspiratory/Expiratory Tidal Volumes

The Analyzer measures tidal volume when performing ventilator measurements. This

option allows the user to choose between the expiratory or the inspiratory tidal volume.

The default setting is inspiratory.

Note

This setting applies only when the bi-directional breath-detection mode is

selected.

To change between inspiratory and expiratory tidal volume:
1. From the

Gas Settings

menu, highlight

Bi-Dir Tidal Vol

.

2. Press the

MODIFY

soft key to toggle to the desired value.

3. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

4. Press the

BACK

soft key to return to the

Settings

menu.

Zero Mode

All pressure and differential pressure (flow) sensors in the Analyzer must be periodically

zeroed or calibrated to a zero reference. This zero function can be set to occur

automatically at pre-determined intervals or only when the user requests the function by

pressing the key labeled

ZERO/7

. When a zero occurs, valves inside the device open the

sensors to ambient pressure and measure the zero value to be subtracted from future

measurements.

Note

Because no breath parameters are calculated for any breath in which a

zero occurs, it may be desirable to turn off automatic zeroing for a short

while during a critical test.

To turn automatic zeroing on or off:
1. From the

Gas Settings

menu, highlight

Zero Mode

.

2. Press the

MODIFY

soft key to toggle to the desired value.

3. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

4. Press the

BACK

soft key to return to the

Settings

menu.