System diagnostics, Linearization – Fluke Biomedical VT Plus HF User Manual

Page 49

Advertising
background image

Connection and Setup

Utilities

2

2-21

4. Press the

ZERO

soft key.

5. Press

OK

to start 21 % O

2

calibration. Allow the full 2-minute period. Do not press

SKIP

.

6. Verify that the O2 reading is 20.9

±

0.1 %.

7. Press the

CAL

soft key.

8. Apply 10 lpm of dry 100 % oxygen to the high-flow inlet or exhaust.
9. Press

OK

to start 100 % O2 calibration. Allow the full 2-minute period. Do not press

SKIP

.

10. Verify that the O

2

reading is 100

±

0.1 %.

11. Press the

BACK

soft key to return to the

Utilities

menu.

Note

If the oxygen sensor cannot be calibrated, it may need to be replaced. Refer

to Maintenance, Service, and Calibration: Oxygen Sensor Replacement.

System Diagnostics

These functions are for the technical service personnel to use while repairing the

Analyzer. They include

Zero All

,

AD Diagnostics

, and

Signal Diagnostics

.

Linearization

This option is most useful to technical service personnel repairing the Analyzer. There is

no input, but the calibration data for each sensor is stored and displayed, so a checksum

test can be performed to make sure the sensor data is still valid.
To check linearization data:
1. From the

Utilities

menu, highlight

Linearization

. The

Linearizaion Info

screen

displays:

fec036.bmp

2. Press the

BACK

soft key to return to the

Utilities

menu.