Special functions, Trend test – Fluke Biomedical VT Plus HF User Manual

Page 71

Advertising
background image

Operation

Special Functions

3

3-21

The Analyzer has conversion factors for gas, volume, and flow measurements, as shown

in Table 3-5.

Table 3-5. Available Conversion Factors

Conversion Factor

Description

ATP

Conversion for gas flow and volume to ambient temperature
and pressure

STPD0

Conversion for gas flow and volume to a standard temperature
of 0 °C and a standard barometric pressure of 760 mmHg

STPD21

Conversion for gas flow and volume to a standard temperature
of 21 °C and a standard barometric pressure of 760 mmHg

BTPS

Conversion for gas flow and volume to body temperature of 37
°C with 100 % relative humidity

Special Functions

Trend Test

The Analyzer provides an automated Trend test that checks a single breath parameter

over an extended period of time to see if it deviates from limits set by the user. The limits

are set as a percent variation from the initial measured value.
To access the Trend test screen, use either of the following procedures:
1. Press the

FULL/5

key to display the breath parameters.

2. Press the soft key

TREND

soft key.

Or
1. Press the

MORE/4

key.

2. Use the S and T soft keys to highlight

Trend

Test

.

3. Press the soft key

ENTER

soft key

.

The

Trend Test Setup

screen displays.

fec057.bmp

To set test parameters: