Fluke Biomedical VT Plus HF User Manual

Page 128

Advertising
background image

VT Plus HF

Operators Manual

A-2

If fluid enters into the pressure measurements connections of the Analyzer, it is important

to evacuate the fluid before storage and before any pressure measurements are made. The

fluid can be evacuated by connecting a 100 cc syringe directly to the pressure

measurement connection and then pulling back the syringe to create a vacuum.

Internal Pressure Measurements

The Analyzer also has several internal pressure measurements. These include barometric

pressure, differential pressure for the flow measurement, and airway pressure.
Airway pressure is measured internally in the tube for the flow sensors. This is done for

convenience for the user while measuring ventilator parameters. Having a pressure

measurement for airway pressure in addition to the external low-pressure measurement

ensures that the pressure is measured at a consistent location and eliminates the need for

cumbersome adapters for placing pressure taps in the airway circuit.
When the Analyzer is set to measure from the high-flow port, the airway pressure is

measured from the high-flow port. Likewise, when the low-flow range is selected, the

airway pressure is measured from the low-flow port. In both cases, the airway pressure

measurement is made on the exhaust side of the flow sensor.

Zeroing the Pressure Signals

The zero reading (offset) of the pressure measurements may drift slightly with time and

temperature. A zeroing function is provided for the user to zero the offset drift. Typically,

this is done when a non-zero reading occurs when there is zero applied pressure.

However, it is good practice to zero the respective signal before any measurement is

taken.
Although a five-minute warm-up period is built into the Analyzer power-up cycle, a 40-

minute warm-up period is suggested for maximum measurement accuracy and stability.

The offset drift is most significant while the sensors for the pressure measurements are

warming up. If the warm-up period is skipped, be aware that the offset drift may be

significant for the first 40 minutes of operation. If the Analyzer has been through an

extreme temperature change, additional time may be required for the warm-up. If the

Analyzer has only been turned off for, say a minute, the warm-up period may be skipped.
The zeroing function for pressure acts as a tare function. As such, the applied pressure

when the signal is zeroed is saved as the new zero reading. The applied pressure during

zero would then be read as zero. Measured pressures would have an offset error equal to

the applied pressure when the Analyzer was zeroed.

Airway Zeroing, A Special Case

The airway pressure may be zeroed manually using the

ZERO/7

key while the airway

pressure screen is active. The airway pressure is also zeroed automatically when the flow

zero occurs. Note that the flow zero may either be performed manually or in an auto zero

mode.

Barometric Pressure, Another Special Case

The offset drift can be adjusted on the barometric pressure measurement by entering the

correct barometric pressure on the

Setup > Settings

menu. Although typically not

needed, it provides a means for the user to ensure that the barometric pressure

measurement is as accurate as possible.

Advertising