Loop screen – Fluke Biomedical VT Plus HF User Manual
Page 95

Remote Operation
Main Screens and Tests
4
4-11
Loop Screen
The Loop, or Loop Plots, screen, shown in Figure 4-6, displays plots of Flow vs. Volume
and Volume vs. Airway Pressure.
fec088.bmp
Figure 4-6. Loop Screen
These plots can be independently re-scaled either automatically or manually.
Controls located on the right-hand side of the screen are used to control the appearance of
the loop plots:
Clear Volume
– allows zeroing of the volume when breath detect has been turned off
Clear Plots
– clears both of the loop plots
Clear Both
– performs both Clear Volume and Clear Plot functions
Number of Loops
– allows specification of the number of breaths that are drawn
before the plots are cleared; one to four breaths may be drawn before the plots are
cleared.