Other setup functions, Breath detect, Breath detect threshold – Fluke Biomedical VT Plus HF User Manual

Page 115: Gas settings

Advertising
background image

Remote Operation

Other Setup Functions

4

4-31

Other Setup Functions

VT for Windows allows selection of settings in the

Setup

menu.

Breath Detect

The Analyzer has a built-in breath detection algorithm. The algorithm determines the

breath phase (inspiration, expiration, inspiratory pause, single phase bi-directional) This

algorithm is designed to reject signal noise and other artifact when determining the

phases of a breath.
To change the breath detect mode:
1. From the

Setup

menu, select

Breath Detect

.

2. Click the preferred option.

Breath Detect Threshold

The Analyzer breath detection algorithm uses a flow rate threshold that the flow must

cross to trigger a change in the breath phase. Higher flow rate thresholds yield less

sensitive breath detection. Lower thresholds yield more sensitive breath detection.
To enter a new breath detect threshold:
1. From the

Setup

menu, select

Breath Detect

; then select

Threshold

. The

following dialog box displays:

fec115.bmp

2. Enter a new value and click

OK

.

Gas Settings

Through the

Gas Settings

option, the Analyzer can compensate for the effects of gas

temperature, viscosity, molecular weight, and relative humidity, on flow measurement, as

well as correct flow calculations to one of several standard conditions. For more

information on the gas settings, refer to System Setup.
To change the gas settings:
1. From the

Setup

menu, select

GAS SETTINGS

. The following dialog box appears,

allowing changes to be made: