Stacked volume test – Fluke Biomedical VT Plus HF User Manual

Page 77

Advertising
background image

Operation

Special Functions

3

3-27

Stacked Volume Test

The Stacked Volume test measures the volume of multiple breaths to determine the

accuracy of the ventilator in accumulating volume.
To access the Stacked Volume test:
1. Press the

MORE/4

key.

2. Use the S and T soft keys to highlight

Stacked Volume Test

.

3. Press the

ENTER

soft key.

4. Press the

MODIFY

soft key.

5. Use the keypad to enter the number of breaths for the test and press the

ENTER

soft

key.

To run the test:
1. Press the

START

or

AUTO

soft key to begin the test. The display changes to show

the progress of the test:

fec067.bmp

The values shown are described below:

Total Volume

– The total volume of all breaths together

Last Volume

– The volume of the last breath taken

Avg. Volume

– The average volume of all breaths taken

Num. Of Breaths

– The total number of breaths taken

Total # Breaths

– The total number of breaths to be taken

Note

If a stacked volume test is started with the

START

soft key, the test

accumulates the requested number of breaths and ends. If the test is started

using the

AUTO

soft key, the test automatically restarts after the requested

number of breaths have accumulated.

2. When the test is complete, press the

STACK

soft key to start a new stacked volume

test or press the

EXIT

soft key to exit.