Barometric pressure, High-pressure measurement, Oxygen concentration – Fluke Biomedical VT Plus HF User Manual

Page 54: Calculated breath parameters

Advertising
background image

VT Plus HF

Operators Manual

3-4

Barometric Pressure

A barometric pressure measurement is provided by the Analyzer. The barometer can read

absolute pressures from 8 to 18 PSIA. The barometer is also used in the automatic ATP,

BTPS, and STPD conversions for flow and volume measurements. The barometric

pressure signal can be fine tuned using the procedure described in System Setup.

High-Pressure Measurement

The high-pressure measurement is made from a dual-port connection consisting of a (+)

positive and a (-) negative pressure port. The differential pressure range is ±100 psi. This

pressure measurement can be used for any pressure differential or gauge pressure in the

given range. The transducer is capable of measuring fluid pressure on the (+) positive

port as indicated on the label.

Oxygen Concentration

Oxygen concentration is measured through the high-flow port of the system on the rear

panel bulkhead fitting. An integrated sensor measures the oxygen percent of the gas in

the high flow channel of the Analyzer. The range for this sensor is 0 % to 100 %. The

oxygen cell is mounted inside the enclosure on the rear bulkhead for the high-flow

circuit. The fuel cell must be replaced approximately once a year. Oxygen readings are

compensated for airway pressure.

Calculated Breath Parameters

From the primary flow and pressure measurements, the Analyzer is designed to calculate

breath rate, % leak, and base flow. A breath detection algorithm determines the various

phases of a ventilator breath and calculates the parameters listed in Table 3-1.

Advertising