VT Plus HF

Operators Manual

4-32

fec116.bmp

2. Enter changes as desired.
3. If necessary, click the

Advanced

button. The

Gas Settings Date Entry

dialog

box displays.

fec117.bmp

4. Make appropriate entries to adjust temperature, viscosity, molecular weight, and

relative humidity for both inhaled and exhaled gases. Ambient temperature can be

adjusted, and the flow can be adjusted to one of six standard modes (ATP, BTPS,

STPD), as shown below:

5. If the viscosity or molecular weight of the gas being used is unknown, click the

Gas

Calculator

button. The

Gas Calculator

dialog box displays: