Maintenance, Service, and Calibration

Maintenance

5

5-3

Maintenance

The Analyzer requires little maintenance or special care; however, it is a calibrated

measuring instrument and should be treated as such. The optional carry case is

recommended for storage. It is further recommended that the storage environment be free

from vibration.

Avoiding Damage

Do not drop the instrument or subject it to any mechanical abuse that could cause a shift

in the calibrated settings.

W

Caution

To avoid

damage to the Analyzer or adverse affects on its

performance, d

o not expose the system to temperature

extremes. Ambient temperatures should remain between 10 °C
and 40 °C.

Cleaning

Clean the exterior of the Analyzer occasionally with a cloth dampened with a mild

detergent solution. A solution of 70 % isopropyl alcohol may be used to remove stains

and clean the system. No other solvents are recommended.

W

Caution

To avoid damage to the Analyzer or adverse affects on its
performance, do not spray liquid directly on or immerse the
unit.

Oxygen Sensor Replacement

The Analyzer uses a galvanic oxygen sensor that is under warranty for 12 months. The

oxygen sensor cell must be replaced when the oxygen calibration cannot be successfully

completed. The oxygen sensor may operate for longer than 12 months depending on use.
To replace the oxygen sensor:
1. Disconnect the power cord before opening the oxygen sensor door.
2. Remove the three screws in the oxygen sensor door on the back of the Analyzer. The

oxygen sensor is a round plastic cylinder about one inch in diameter with wires

protruding from the back of the sensor.

3. Disconnect the in-line electrical connector on the sensor wire.
4. Un-screw the old sensor and remove it.
5. Screw in the new sensor (Fluke part #2138514).
6. Re-connect the sensor wire.
7. Replace the oxygen sensor door.
8. Re-calibrate the oxygen sensor following the instructions under Connection and

Setup: Utilities: Oxygen Sensor Calibration.