Measured signals, High flow, Low flow – Fluke Biomedical VT Plus HF User Manual

Page 53: Airway pressure, Low pressure

Advertising
background image

Operation

Measured Signals

3

3-3

Measured Signals

The Analyzer measures the following signals:

High flow

Low flow

Airway pressure

Low pressure

Barometric pressure

High Pressure

Oxygen concentration

Calculated breath parameters

High Flow

The Analyzer has a high-flow (

±

300 LPM), bi-directional flow measurement. Flow

measurements can be either static flows (no breath variations) or ventilator waveforms

(i.e., both an expiratory and an inspiratory phase). The flow measurement is made using a

differential pressure drop across a resistive screen mesh. This flow measurement can be

used for pediatric/adult ventilators or for determining the performance of many types of

flow meters. Valves switch between the low-flow and the high-flow measurements

depending on the range selected in the display flow screens.

Low Flow

The Analyzer has a low-flow (

±

25 LPM), bi-directional flow measurement. The flow

measurements can be either static (no breath variations) or ventilator waveforms (i.e.,

both an expiratory and an inspiratory phase). Flow measurements are made using a

differential pressure drop across a resistive screen mesh. The low-flow signal can be used

for infant/neonatal ventilators or for determining the performance of many other types of

flow meter. Valves switch between the low-flow and the high-flow measurements

depending on the range selected in the display flow screens.

Airway Pressure

The airway pressure is measured from a tap off of the proximal (near the exhaust port) of

the flow sensor housings. This tap is connected to a

±

120 cmH

2

O pressure sensor that

generates the airway pressure signal. Valves switch between the low-flow and the high-

flow measurements. When the low-flow port is selected, the airway pressure is measured

off the low-flow port. Likewise, when the high-flow port is selected the airway pressure

is measured off the high-flow port. Use the

RANGE

soft key from any of the flow

screens to select which flow port (high or low) airway pressure is measured from.

Low Pressure

This is a dual-port connection consisting of a (+) positive and a (-) negative pressure port.

The differential pressure range is ±500 mmHg. This pressure measurement can be used

for any pressure differential or gauge pressure in the given range. The transducer is

capable of measuring fluid pressure on the (+) positive port as indicated on the label.

Advertising