Fluke Biomedical VT Plus HF User Manual

Page 80

Advertising
background image

VT Plus HF

Operators Manual

3-30

The soft keys for the RT200 screen are:

SETUP RS232

– displays a screen for setting the baud rate, number of data bits, and the

terminator character:

fec071.bmp

These settings affect only the RT200 emulation mode and not the other serial

communication modes. The default settings are 9600 baud, 7 bits, and a termination

character of carriage return only.

PEAK

– toggles the RT200 peak-detect mode on and off, for flow and pressure signals;

when in peak-detect mode, the peak of the current signal (flow or pressure) is shown for

each breath

RESET

– resets the RT200 mode, turning off any active functions, and resetting the

special functions

FCTN

– displays a menu that allows the user to start a RT200 function and enter the

special function values associated with that RT200 function. The menu is function

sensitive, and the options in the menu correlate to the RT200 special functions 00, 01, F1,

and F2. The values set in this menu include those shown below:

fec070.bmp