Remote Operation

Main Screens and Tests

4

4-19

Running the Stacked Volume Test

To run the stacked volume test:
1. Click the

START

button to begin testing. The screen changes, as shown in Figure

4-14:

fec096.bmp

Figure 4-14. Stacked Volume Screen: Testing Underway

When the stacked volume test begins, the Analyzer switches into Inspiratory Only

breath-detect mode. At the end of the test, the breath-detect mode is restored to its

previous state. The stacked volume test reports five values to the user; they are

updated every time a breath is detected:

Total Volume

–the total of the tidal volumes for the breaths tested

Last Volume

–the tidal volume for the last breath

Average Volume

– the average tidal volume over the breaths that have been tested and

calculated

Number of Breaths

– the number entered at the start of the test; it does not change during

a test.

Breath Count

–the current number of breaths included in the test; if tidal volume can't be

calculated for a breath, that breath and volume are not included in the test.