Full test screen, Trend test – Fluke Biomedical VT Plus HF User Manual
Page 96
![background image](/files/785361/content/doc096.png)
VT Plus HF
Operators Manual
4-12
Full Test Screen
The Full-Test screen, shown in Figure 4-7, displays the calculated parameters for the last
breath.
fec089.bmp
Figure 4-7. Full Test Screen
These parameters are only updated when a new breath occurs. A ***** value means that
value is invalid or could not be calculated. Depending on the breath-detect mode, some
parameters can't be calculated.
Trend Test
The Trend test is used for monitoring a single-breath parameter over an extended period
of time. Each time a breath occurs, the chosen breath parameter is compared to an initial
value. If it exceeds the percent limit given for the test, an incident is reported. Following
the test, a report may be printed detailing all of the incidents that occurred during the test.