Assist test – Fluke Biomedical VT Plus HF User Manual

Page 81

Advertising
background image

Operation

Special Functions

3

3-31

Assist Test

The Assist test determines the sensitivity of the ventilator in patient assist mode and helps

to calibrate the full range of assist pressures under which the ventilator can operate.

During an assisted breath, the airway pressure drops during the inspiration due to the

spontaneous breath. The assist is triggered by a negative pressure (relative to the end-

expiratory pressure) during the expiratory hold period. See Figure 3-2.

AIR

W

AY

PRESSURE

ASSISTED

BREATH

TIME

N

ORMAL

BREATH

O

MI

N

IMUM AIR

W

AY

PRESSURE

EEP

fec072.eps

Figure 3-2. The Assist Test

To perform an Assist test:
1. Press the

FULL/5

key. The

Full Test

screen displays.

2. Connect the ventilator breathing circuit to the Analyzer in the bidirectional

measurement configuration as shown in Figure 3-3.

O

u

tlet

Inlet

V

entilator

Inspiratory Hose

Expiratory Hose

Analyzer

TEST
LU

N

G

fec002.eps

Figure 3-3. Ventilator Breathing Circuit Setup for Bi-Directional Measurements

3. Adjust the ventilator controls so that the ventilator is triggered by an effort (negative

pressure) at the lung during the expiration hold period of the ventilator's breath cycle.

Note

Use the Synchronized Intermittent Mandatory Ventilation (SIMV) mode or

Intermittent Demand Ventilation (IDV) mode if the ventilator provides it.

4. Check the control setting by applying upward force at the back of the test lung and

observing if the ventilator triggers as desired.

5. Set a low breath rate to allow the distinction to be made between a triggered assist

breath and a mandatory breath.