Volume screen – Fluke Biomedical VT Plus HF User Manual

Page 66

Advertising
background image

VT Plus HF

Operators Manual

3-16

Volume Screen

The Volume screen displays the signal and statistics for the volume signal.

fec053.bmp

The volume signal is the integral of the flow signal measured in either the high or low

flow ports. The flow port that is integrated is selected by selecting the high- or low-flow

range from within the flow screen.
Flow Zero

Press the

FLOW ZERO

soft key to zero the flow reading.

Units Press

the

UNITS

soft key to change the units of measure for the volume

screen. Possible units of measure for this screen include:

l

Liters

ml

Milliliters

CF

Cubic

Feet

Re-scale Press

the

RESCALE

soft key to change the scale on the plot of data to a

more appropriate window of values.

Note

RESCALE

uses the current min/max values to

determine plot limits.

Clear Press

the

CLEAR

soft key to reset the statistics and re-start the plot.

The user can select the breath parameters displayed on this screen by following the

instructions under Selecting Parameters.

Note

The volume signal is integrated from either the high- or low-flow signals.

To select which flow signal is used to calculate volume:
1. Press the

FLOW/0

key.

2. Press the

RANGE

soft key until the screen corresponding to the desired flow signal

(high or low) displays.