Outputs – Metrohm IC Net 2.3 User Manual

Page 101

Advertising
background image

6.5 761 Compact IC

IC Net 2.3

91

Cell constant

Cell constant of the conductivity cell for cor-
rect display of the absolute conductivity.
Enter the cell constant printed on the detec-
tor block into this field.
Range:

0.1

...

99.9 /cm

For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.

<Break times>

Possibility for changing the break times for
injection valve

Valve

and suppressor mod-

ule

Suppressor

.

These values should only be changed af-
ter consulting the Metrohm Service.


Outputs

The

Outputs

tab of the hardware settings window defines remote

output signals to be set automatically if specific events occur.

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the "IN-
JECT" position.

Fill

Switching of the injection valve to the "FILL"
position.

Suppressor step

Switching of the suppressor module to the
next position.