System settings – Metrohm IC Net 2.3 User Manual

Page 270

Advertising
background image

6 Devices

IC Net 2.3

260

System settings

On the tab

System settings

break times for the injection valve and

flow rate corrections for the high-pressure pump are defined.

Valve

Injection

valve.

Break time, ms

Modification of the break time for the injec-
tion valve.
Entry range:

1

...

99 ms

Pump

High pressure pump.

Flow correction

Factor for correction of the difference bet-
ween displayed and actual flow rate of the
high-pressure pump.
Range:

0.9 ... 1.25

The correction factor is determined as fol-
lows by measuring the actual flow rate us-
ing a graduated cylinder:

Flow correction

=

Displayed flow rate
Measured flow rate

The product of parameter

Flow correction

and parameter

Flow

(on the tab

Controls

)

must not exceed 2.75 mL/min.

Ch1/Ch2 threshold

Limit for measuring channel ratio Ch1/Ch2.

Threshold, mAU

As long as the absorption value(s) from

channel 1

,

channel 2

or

both

lie beneath the

defined

Threshold

(limit) the value for the

Ch1/Ch2 ratio will not be calculated.

Analog Outputs

Settings for analog signal output.

Range, mAU

The defined value is equivalent to 1 V for
the analog signal output.
Entry range:

0.1

...

2000.0 mAU