Integration events for peak detection – Metrohm IC Net 2.3 User Manual

Page 295

Advertising
background image

7.4 Integration

IC Net 2.3

285

If an integration event is added or modified, the

Edit integration

event

window is opened where the following parameters can be

set:

Time

Start time for integration event in min.

Event

List box with integration events to be se-
lected (see below).

Value

Parameter value for selected integration
event. This field appears for events that
demand input of an additional parameter.

Integration events for peak detection

Disable detection

Set this mode to stop detection of new
peaks. If a peak is available at the moment
of the event it is either finished (downslope
peaks) or rejected (upslope peaks).

Enable detection

Clear

Disable detection

mode.

Enable negative peaks

Enable detection of negative peaks. In

some cases this mode may result in insta-
bility of detection algorithm.

Disable negative peaks

Disable detection of negative peaks. This

event does not influence negative peaks
which already started.

Disable peak reject

Set this mode, if a small peak should not be
rejected because of its flat apex.

Enable peak reject

Clear

Disable peak reject

mode.

Advertising