Metrohm IC Net 2.3 User Manual

Page 362

Advertising
background image

10 Batch reprocessing

IC Net 2.3

352

switched on) by updating the concentration
table.

Default scheme

Default setting for recalibration reprocess-
ing. The two options

Apply final calibration...

and

Forget calibration points...

are switched

on. A new calibration is performed with the
calibration runs and the resulting new cali-
bration parameters (component table, con-
centration table and calibration curve) are
applied to all sample runs.

Apply final calibration to all reprocessed files

Apply the updated calibration to all calibra-
tion and sample runs. If this option is
switched off, the calibration stored in the
first chromatogram is used for all other
chromatograms.

Forget calibration points before reprocessing

Forget old calibration points of the calibra-
tion curve and perform a new calibration
using all calibration runs of the batch re-
processing queue. In this case each cali-
bration chromatogram adds a new point to
the calibration curve. If this option is
switched off, the calibration curve stored in
the first chromatogram remains active.
Each further calibration chromatogram in
the batch reprocessing queue adds a new
point to this calibration curve.

Update retention time on runs:

Update the retention times in the peak table
after the run. Tool to analyse difficult sam-
ples. Compensates a drift of the retention
time to enable correct peak identification
and integration.

Optimize your analysis first to get stable
measuring conditions without a retention
time drift. In any case it is better to have
stable conditions, use this feature only as
a last possibility.

Calibration

If checked, the retention times will be up-
dated in the peak table after every calibra-
tion.

All

If checked, the retention times will be up-
dated in the peak table after every determi-
nation.

Recalculate only

Enable recalculation of chromatograms
with values for

Volume

,

Dilution

,

Amount

and

Advertising