Display – Metrohm IC Net 2.3 User Manual

Page 86

Advertising
background image

6 Devices

IC Net 2.3

76

Display

The

Display

tab of the

819 IC Detector

window is only available for a

connected system. It displays the current measurement values
and parameters settings of the 819 IC Detector.

Conductivity

Live display of measured conductivity
(auto-zero value) in

µ

S/cm. This field is

available for the

WATCH WINDOW

. The

watch window color is changed to the

Out

of range

color if the measured value is more

than 80% outside the

Full Scale

range

.

abs. Cond.

Live display of absolute conductivity value
in

µ

S/cm. This field is available for the

WATCH WINDOW

. The watch window color is

changed to the

Out of range

color if the

measured value is bigger or equal to 110%
of

Range.

Ion

Display of ion settings:

Anion

:

polarity = +, temperature coeff. = 2.5

Cation

:

polarity = –, temperature coeff. = 1.5

Range

Display of measuring range in

µ

S/cm.

Full

scale

Display of full scale range in

µ

S/cm.

Cell

const

. Display

of cell constant of the conductivity

cell in /cm.

Thermostat

Display of temperature within in the detec-

tor block.

Offset

Display of offset of the zero point.