Metrohm IC Net 2.3 User Manual

Page 147

Advertising
background image

6.12 Metrohm 709 Solvent Delivery Unit

IC Net 2.3

137

Program table

Number

Row number in the program table (read-
only).

Time

Time at which the flow and/or gradient is
changed.
Entry range:

0.0 ... 99999 min

Flow

Overall flow of pump control system in
mL/min.
Entry range:

0.01 ... 20.00 mL/min

(depending

on number of pumps)

%B

Individual flow rate for pump B in percent of
total flow.
Entry range:

0 ... 100 %

%C

Individual flow rate for pump C in percent of
total flow.
Entry range:

0 ... 100 %

%D

Individual flow rate for pump D in percent of
total flow.
Entry range:

0 ... 100 %

Enabled

Enable time program start (a disabled pro-
gram is not started).

Show flow column

Show

Flow

column in the program table.

<Add>

Add a new line to the time program.

<Delete>

Delete the selected line from the time pro-
gram.

<Start>

Start all pump drives.