6.11 818 IC Pump

IC Net 2.3

131

used. If the pump exceeds the preset limit
value during operation, it is switched off
immediately. At the same time, the

Pressure

field color in the 818 settings window and in
the

WATCH WINDOW

is changed to the

value set for

Out of range

.

The

set

minimum limit value should lie far

enough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolu-
tions of the pump cam due to leaks or inter-
rupted inflow of the eluent, the pump is
automatically switched off. At the same
time, the

Pressure

field color in the 818 set-

tings window and in the

WATCH WINDOW

is

changed to the value set for

Out of range

.

<Set>

Send current parameters immediately to the
818 IC Pump (only available if pump is run-
ning). Parameters are not stored in the sys-
tem file (

*.smt

) as long as the file is not

saved.

<Start>

Set current parameters and start 818 IC
Pump drive.

<Stop>

Stop 818 IC Pump drive.

Time program

The

Program

tab of the

818 IC Pump

window a user-defined time

program can be entered. This program is started automatically as
defined in the

Start mode

window (see section 4.4.3) either at the

moment the determination is started (

Start with determination

) or at

the moment the sample is injected (

Start with inject

).