Pulse mode – Metrohm IC Net 2.3 User Manual

Page 117

Advertising
background image

6.7 871 Advanced Bioscan

IC Net 2.3

107

Startup values

E

Electric potential applied to the amperomet-
ric detector cell in V
Range:

-2 ... +2 V

.

Range

Selection of measuring range:

pA

10 ... 500 pA

.

nA

1 ... 500 nA

.

µA

1 ... 200 µA

.

Filter

Electronic damping of the analog output
signal.
Selection:

off, 0.2, 0.5, 1, 2, 5 s

Offset

Offset of the zero point for the analog out-
put signal.
Selection:

-50 ... +50 %

(in steps of 10 %)

Data rate

Setting of the data acquisition rate.
Selection:

-1, 2, 5, 10 Hz

Polarity

Set sign:

+

or

-

Actual

E

Display of current cell potential in V

Range

Display of current measuring range.

Filter

Display of current damping constant.

Offset

Display of current offset.

Data rate

Display of current data rate.

Polarity

Display of current polarity.

<Set>

Send current parameters immediately to the
871 Advanced Bioscan. Parameters are not
stored in the system file (

*.smt

) as long as

the file is not saved.

Pulse mode

This subtab of the

Method parameters

tab in the

871 Advanced Bio-

scan

window contains the method parameters for the Pulse

mode. These values are sent and applied to the 871 Advanced
Bioscan by

Start determination

and

Startup hardware

, or if the val-

ues are sent manually with

<Set>

.

Startup values

E1, E2, E3

Definition of three different working poten-
tials, which are applied to the amperometric
detector cell cyclically. The set potentials
are graphically displayed if the

Method

op-

tion is enabled above the graphic window.
Range:

-2 ... +2 V

.

t1, t2, t3

Time duration to apply the working poten-
tial.
Range:

0.1 ... 2 s

(t1)

, 0.01 … 2 s

(t2,t3)