Time program – Metrohm IC Net 2.3 User Manual

Page 131

Advertising
background image

6.9 820 IC Separation Center

IC Net 2.3

121

Valve A

fill/inject

Display of current valve position.*

<FILL>

Switch valve A to "fill" position.

<INJECT>

Switch valve A to "inject" position.

Valve B

fill/inject

Display of current valve position.*

<FILL>

Switch valve B to "fill" position.

<INJECT>

Switch valve B to "inject" position.

Suppressor

in position

Display of current suppressor status.*

<STEP>

Switch suppressor module to next position.

Auto Step

*

fill, inject, off

Display of current auto step status.

Entry field

Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection:

fill, inject, off

* only available for

820 IC Separation Center

connected to 819

Time program

The

Program

tab is only available for a

820.0XX0 Separation Center

connected to a 830 IC Interface. The

820 IC Separation Center

that is connected to a 819 IC Detector is controlled by the time
program of the detector.
On the

Program

tab of the

820.0XX0 Separation Center

window a

user-defined time program can be entered. This program is
started automatically as defined in the

Start mode

window (see

section 4.4.3) either at the moment the determination is started
(

Start with determination

) or at the moment the sample is injected

(

Start with inject

).