List of program instructions, Links – Metrohm IC Net 2.3 User Manual

Page 165

Advertising
background image

6.16 793 Sample Prep Module

IC Net 2.3

155

First column

Time at which program instruction is ap-
plied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.

Second column

Program instruction (see below).

Third column

Parameter for program instruction (see be-
low).

ENABLED

Enable program start (a disabled program
is not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are
displayed if program is wrong).


List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

page:

Instruction Parameter

entry Meaning

Program RESET,

END

Set program flag. The

END

flag

can be used to end a program,

especially if the program time

should be longer than the chro-

matogram

duration

. Additional

steps after this flag are not al-

lowed. The

RESET

flag is used to

reset the parameters to the sys-

tem startup values.

Pump on,

off

Switch on or off the pump drive.

Step

Rotate sample preparation mod-

ule to the next position.

Links

The

Links

tab of the

793 Sample Prep Module

is used for Event line

selection.