Metrohm IC Net 2.3 User Manual

Page 262

Advertising
background image

6 Devices

IC Net 2.3

252

<Off>

Turn VIS lamp off.

<Zero>

Records a new zero spectrum.
The zero spectrum is a reference spectrum.
It should be recorded at a time at which
there is no sample in the measuring cell. It
is used for calculating the absorption spec-
trum.

<Spectrum>

Opens

the

SPECTRUM

window (see Section

6.28.5).

Wavelength, nm

Sets the wavelengths at which the chroma-
tograms are to be recorded (for max. three
measuring channels).

Bunching, nm

Sets the bunching of the wavelengths de-
fined above.

Channel 1, nm

Measuring channel 1.

Channel 2, nm

Measuring channel 2.

Channel 3, nm

Measuring channel 3.


The relationship between Wavelength and Bunching is ex-
plained by using an example:

Set values: Wavelength = 500 nm, Bunching = 21 nm
An upper and a lower limit for the measured wavelengths of
the corresponding measuring channel are calculated from
these values. For these calculations the following equation
applies: subtract "1" from the

Bunching

value, divide the new

value by 2, and subtract/add the resulting value from/to the

Wavelength

value.

Result: wavelengths between 490 nm (lower limit) and 510 nm
(upper limit) will be included in the chromatogram.