Metrohm IC Net 2.3 User Manual

Page 256

Advertising
background image

6 Devices

IC Net 2.3

246

The

Hardware

tab of the hardware settings window defines ge-

neral parameters which are set automatically at power on of the
861 Advanced Compact IC.

Remote lines after power on

The remote output lines 1

8 are set to this

values after power on or a manual stop with

Shutdown hardware

.

Selection:

0

,

1

Thermostat

Operating temperature of the conductivity
cell.
Selection:

25

,

30

,

35

,

40

,

45 °C

,

off

Thermostatting functions only if the ambient tempera-
ture is at least 5 °C lower than the operating tempera-
ture. It normally takes 30

60 min after power on until

a temperature stability of

±

0.01 °C is attained.

Flow correction

Factor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range:

0.9

...

1.09

The flow correction is determined by meas-
urement of the actual flow rate with the aid
of a measuring cylinder as follows:

Flow correction

=

Displayed flow rate
Measured flow rate

Cell constant

Cell constant of the conductivity cell for cor-
rect display of the absolute conductivity.
Enter the cell constant printed on the detec-
tor block into this field.
Range:

0.1

...

1000.0 /cm