5 spectrum window – Metrohm IC Net 2.3 User Manual

Page 272

Advertising
background image

6 Devices

IC Net 2.3

262

Operating hours

Operating hours of the lamps.

UV lamp, h

Display of the operating hours of the UV
lamp.

<Reset UV lamp>

Set

operating hours of the UV lamp

to 0. Click

it after replacing the UV lamp.

VIS lamp, h

Display of the operating hours of the VIS
lamp.

<Reset VIS lamp>

Set

operating hours of the VIS lamp

to 0. Click

it after replacing the VIS lamp.

6.28.5 Spectrum

window

The

SPECTRUM WINDOW

can be opened with a right-hand mouse-click on

the 844 icon and then selecting the menu item

Spectrum window

, or via the

button on the subtab

Manual/Detector

.

A selection can be made from three different spectra, the absorption

spectrum, intensity spectrum and the zero spectrum. Please note that each

time you click on

Absorption

,

Intensity

or

Zero

the function of the

<Update>

button will be triggered.

<Update>

The 844 measures the current intensity spectrum and

sends it together with the last zero spectrum to have

been recorded to <IC Net>. The absorption spec-

trum is recalculated from these two spectra.

Absorption

The absorption (with the units mAU) is given on the y-

axis as a function of the wavelength (x-axis). The ab-

sorption values are calculated for all measured wave-

lengths and shown a s spectrum. The calculation for-

mula for the absorption value is shown here as an ex-