Alarm stops – Metrohm IC Net 2.3 User Manual

Page 258

Advertising
background image

6 Devices

IC Net 2.3

248

1

(line on, active, 0 V)

p

(pulse output)

*

(leave line in current status)

PowerOn values

Display of the power on startup values for
remote output lines set on the

Hardware

tab.

Pulse length

Length of a pulse in ms.

Alarm stops

The

Alarm stops

tab of the hardware settings window defines the

events for which the 861 Advanced Compact IC is stopped im-
mediately. At an alarm stop, high-pressure pump and peristaltic
pump are stopped immediately, the running determination and
the active sample queue are also stopped.

Event

Events for alarm stop:

Pump stop

Pump stopped because pressure limits are
exceeded.

Hardware error

Hardware error detected at the 861 Ad-
vanced Compact IC (high-pressure pump,
injection valve, or suppressor not working
correctly).

Leak

Leak detector has detected solvent in the
instruments interior. This information is also
stored in the instrument itself, so that it is
stopped automatically even without con-
nection to the PC.

Power on

Temporary power failure at the 861 Ad-
vanced Compact IC hardware.