6.10.3 709 IC Pump window

709 icon / Open

The

709 IC Pump

window for parameter settings and manual con-

trol is opened by selecting this menu option with the right mouse
button or by double-clicking the

709 icon

in the

SYSTEM

window. It

consists of the three tabs

Settings

,

Links

and

Advanced

.

Settings

The

Settings

tab of the

709 IC Pump

window is only available for a

connected system.

Pressure

Live display of current pressure in MPa.
This field is also available for the

WATCH

WINDOW

.

Flow

Live display of current flow in mL/min. This
field is also available for the

WATCH WIN-

DOW

.

New flow

A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range:

0.01 ... 5.00 mL/min

Min/Max pressure

The

minimum and maximum pressure limit

for the 709 IC Pump can be set in steps of
0.1 MPa.
Entry range:

0.0 ... 50.0 MPa

.

The

set

maximum limit value should lie be-

tween 5 to 10 MPa above the particular
operating pressure or the maximum admis-
sible operating pressure of the column
used. If the pump exceeds the preset limit