Time program – Metrohm IC Net 2.3 User Manual

Page 181

Advertising
background image

6.21 837 IC Degasser

IC Net 2.3

171

Time program

On the

Program

tab of the

837 IC Degasser

window a user-defined

time program can be entered. This program is started automati-
cally as defined in the

Start mode

window (see 4.4.3) either at the

moment the determination is started (

Start with determination

) or at

the moment the sample is injected (

Start with inject

).

First column

Time at which program instruction is ap-
plied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.

Second column

Program instruction (see below).

Third column

Parameter for program instruction (see be-
low).

Enabled

Enable program start (a disabled program
is not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are
displayed if program is wrong).