Time program – Metrohm IC Net 2.3 User Manual

Page 88

Advertising
background image

6 Devices

IC Net 2.3

78

Custom

independent selection of

polarity

and

tem-

perature coeff.

Polarity

Polarity of the analog output signal:

+

positive polarity (e.g. for anions)

negative polarity (e.g. for cations)

Temperature

coeff.

Converts automatically the conductivity
from the operating temperature of the
measuring cell to the reference temperature
of 20°C.
Selection:

1.5, 2.5 %/°C

Cell

constant

Cell constant of the conductivity cell in /cm.
Entry range:

13 ... 21 /cm

Thermostat

Operating temperature of the conductivity
cell within in the detector block.
Selection:

25, 30, 35, 40, 45 °C

and

off

Offset

Offset of the zero point of the conductivity.
Selection:

0, 10, 50 %fs

Damping

Electronic damping of the analog output
Selection:

on, off

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1

<SET>

Send current parameters immediately to the
819 IC Detector. Parameters are not stored
in the system file (

*.smt

) as long as the file is

not saved.

For detailed information about the parameters of the 819 IC De-
tector see 819/820 Instructions for Use.

Time program

On the

Program

tab of the

819 IC Detector

window a user-defined

time program for the remote output lines of the 819 IC Detector
can be entered. This program is started automatically as defined
in the

Start mode

window (see section 4.4.3) either at the moment

the determination is started (

Start with determination

) or at the

moment the sample is injected (

Start with inject

).