6 Devices

IC Net 2.3

214

End time

Subtab for the end time settings of the
timebase program.

Current

Displays the current

End time

of the Triath-

lon autosampler.

Method value

Manual setting of the end time. The end
time is sent to the Triathlon Autosampler
with

<SET>

.

Entry range:

0:00:00…9:59:59 h:mm:ss

auto time

enabled

Auto setting of the end time is enabled, the
end time is equal to the analysis time (see

Methods programming/Triathlon meth-
ods/Injection/Injection

) if the end of a time-

base program exceeds the analysis time, it
is overruled.

disabled

The end time specified under

Method value

is selected, if a timebase program exceeds
the end time, the programmed events after
the end time are not carried out during the
run.