Remote configuration – Metrohm IC Net 2.3 User Manual

Page 253

Advertising
background image

6.27 861 Advanced Compact IC

IC Net 2.3

243

Instruction Parameter

entry Meaning

Valve Inject,

Fill

Switch injection valve to "INJECT"

or "FILL" position.

FullScale

50, 250, 1000, 5000

µ

S/cm

Set full scale range to the se-

lected value.

ICPump on,

off

Switch on or off the high-pressure

pump.

Flow

0.2 ... 2.5 mL/min

Set flow rate of the high-pressure

pump to the desired value.

Pmax

0.0 ... 35.0 MPa

Set maximum pressure limit for

the high-pressure pump to the de-

sired value.

Pmin

0.0 ... 35.0 MPa

Set minimum pressure limit for

the high-pressure pump to the de-

sired value.

Remote

0, 1, *, p

Set remote output lines 1...8 to

the desired values. For entry of the

first value, enter

1

,

0

,

p

or

*

. For en-

try of the other values, move the

cursor in front of the value to be

changed and enter

1

,

0

,

p

or

*

.

Program END,

RESET

The

END

flag can be used to end a

program, especially if the program

time should be longer than the

chromatogram duration. Additional

steps after this flag are not allowed.

The

RESET

flag is used to reset the

parameters to the system startup

values.

Suppressor

Switch suppressor module to the

next position.

Peristaltic on,

off

Switch on or off the peristaltic

pump.

Remote configuration

On the

Remote configuration

subtab user-defined remote com-

mands can be defined, which can be inserted into a time pro-
gram after being activated with

<Activate>

.