4 processing – Metrohm IC Net 2.3 User Manual

Page 286

Advertising
background image

7 Methods

IC Net 2.3

276

bit smaller and wider after smoothing, but
their area does not change.

Raw data points

Number of data points per chromatogram
(read-only).

Smallest peak has ...

Number of data points per smallest peak

on the chromatogram (read-only). This in-
formation is valuable to perform the

Com-

press

procedure (data compression).

7.3.4 Processing

Tab

Processing

of

Method setup

window for definition of actions

that are performed during the run or when the chromatogram fin-
ishes. Automatic post-run data processing is especially useful for
chromatographic systems that include an autosampler.

Save chromatogram after the run

If checked, autosaving of the chroma-
togram will be performed on data acquisi-
tion ending.

while running save every ... min

Save the chromatogram during the run for
data safety at the time interval set (if

0

is

set, the chromatogram is not saved during
the run).

Update retention time on runs:

Update the retention times in the peak table
after the run. Tool to analyse difficult sam-
ples. Compensates a drift of the retention
time to enable correct peak identification
and integration.