List of program instructions – Metrohm IC Net 2.3 User Manual

Page 90

Advertising
background image

6 Devices

IC Net 2.3

80

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are
displayed if program is wrong).

List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

subpage:

Instruction Parameter entry

Meaning

Remote

0, 1, *

Set remote output lines 1...8 to

the desired values. For entry of

the first value, enter

1

,

0

or

*

. For

entry of the other values, move

the cursor in front of the value to

be changed and enter

1

,

0

or

*

.

Range

100, 200, 5000

µ

S/cm

1, 2, 5, 10 mS/cm

Set measuring range to the se-

lected value.

FS 0.05

µ

S/cm ...

10 mS/cm

Set full scale range to the se-

lected value (depends on

Range

setting).

Zero on,

off

Switch on or off the auto-zero

function.

Polarity +,

Reset the polarity of the output

signal.

Mark

Trigger marking signal.

ValveA fill,

inject

Switch injection valve A to "

inject

"

or "

fill

" position.

ValveB fill,

inject

Switch injection valve B to "

inject

"

or "

fill

" position.

Suppressor

Switch suppressor module to the

next position.

StartDA

Start data acquisition.